Trade Mark No.: 2022609
Bionics Scientific is a Class B dental autoclave manufacturer in India, offering 16, 18 and 23 litre front-loading steam sterilizers for dental clinics, hospitals, veterinary clinics and medical facilities. These dental autoclaves use pre-vacuum air removal, saturated steam sterilization and post-vacuum drying for compatible wrapped, unwrapped, solid, porous and hollow instruments.
The available models operate at 121°C and 134°C and combine programmable cycles with a stainless steel chamber, digital controls, independent pressure monitoring and a double-locking safety system. Bowie-Dick, Helix and vacuum test programs help operators routinely check air removal, steam penetration and vacuum performance according to their facility procedures.
Our Class B sterilizers are designed with reference to EN 13060 requirements for small steam sterilizers. Buyers should confirm the certification, validation documents and configuration supplied with the selected model before ordering.
Three chamber capacities help dental practices select a machine according to daily patient volume, instrument turnaround and available installation space. The 16 litre model suits lower instrument loads, while the 18 and 23 litre models provide additional capacity for growing or multi-chair clinics. Final selection should be based on the dimensions and packaging instructions of the instruments being processed.
| Adopts European Class B standard | Excellent vacuum dry effect |
| Automatic water filling option | Built-in steam generator |
| High quality stainless steel chamber | Meets European Standard EN 13060 |
| User-friendly operation | Digital control panel |
| Descriptive indicators and alarms | Independent pressure gauge |
| Built-in printer | PC connection interface |
| Overheat protection | Double-locking safety system |
| Model | BST/DA-16 | BST/DA-18 | BST/DA-23 | |
|---|---|---|---|---|
| Sterilization Class | European Class B standard Dental Autoclave | |||
| Capacity | 16 liters | 18 liters | 23 litres | |
| Sterilizing temperature | 121°C, 134°C | |||
| Sterilization applications | Compatible wrapped, unwrapped, solid, porous and hollow instruments | |||
| Dry procedure | Vacuum drying | |||
| Display | LED screen | |||
| Test program | Bowie & Dick test, Helix test, Vacuum test | |||
| Safety Features | Safety valve, double-control locking system and self-test alarm | |||
| Water supply system | Two water tanks inside | |||
| Capacity of water tank | 3.5 litre clean-water tank and 5 litre waste-water tank | |||
| Instrument container | 5-layer tray rack with 3 trays | |||
| Voltage | 220V±10%, 50Hz | |||
| Optional Accessories | -Print by optional external mini-printer or download from USB port ( Sterilizing
data) -Extra stainless steel trays |
|||
| Program | Temperature | Pressure (bar) | Holding Time (min) | Total Cycle Time (min) |
|---|---|---|---|---|
| SOLID | 134°C | 2.1 | 4 | 18-30 |
| SOLID | 121°C | 1.1 | 20 | 30-45 |
| WRAPPED | 134°C | 2.1 | 6 | 30-40 |
| WRAPPED | 121°C | 1.1 | 20 | 35-50 |
| TEXTILE | 134°C | 2.1 | 10 | 45-65 |
| TEXTILE | 121°C | 1.1 | 30 | 50-75 |
| PRION | 134°C | 2.1 | 18 | 45-70 |
| PLASTIC | 105°C | 0.2 | 30 | 50-65 |
| B&D TEST | 134°C | 2.1 | 3.5 | 22-35 |
| VACUUM TEST | ------ | ----- | 3.5 | 4-10 |
A Class B cycle first removes air from the chamber and instrument load through vacuum pulses. Saturated steam then contacts the load at the selected temperature for the programmed holding time. After exposure, post-vacuum drying removes residual moisture before the cycle is completed. The four essential steam-sterilization parameters are steam, pressure, temperature and time.
Routine monitoring helps a clinic confirm that the sterilizer is operating as intended. The available test programs include:
Mechanical, chemical and biological monitoring should be performed according to the manufacturer’s instructions and the clinic’s applicable infection-control requirements. For independent technical guidance, see the CDC guidance on steam sterilization.
Tell us your preferred 16, 18 or 23 litre capacity and daily instrument load. Our sales team will share the suitable model specifications, available options and quotation.
Request Price & BrochureA Class B dental autoclave is an advanced steam sterilizer that uses a vacuum system to remove air before sterilization. It is suitable for sterilizing wrapped, hollow, and solid dental instruments.
A Class B autoclave uses pre-vacuum air removal to help steam reach compatible wrapped, porous and hollow dental instruments when the validated cycle and loading instructions are followed.
A dental autoclave sterilizes instruments by using high-pressure saturated steam at high temperatures to destroy bacteria, viruses, fungi, and spores.
Class B autoclaves can sterilize dental handpieces, surgical instruments, mirrors, forceps, scalers, wrapped instruments, and hollow devices.
Cycle duration depends on the selected program, temperature and load. The listed Bionics models provide total cycle times from approximately 18 to 75 minutes.
Dental autoclaves typically operate at 121°C or 134°C, with 134°C being commonly used for rapid and effective sterilization.
Yes, a Class B dental autoclave is specifically designed to sterilize wrapped instruments, allowing them to remain sterile until use.
Testing and maintenance frequency should follow the manufacturer’s instructions, applicable regulations and the clinic’s infection-control protocol.
Common causes include overloading the chamber, improper instrument arrangement, insufficient water, blocked filters, or damaged door seals.