Trade Mark No.: 2022609
Only B type cycles are considered safe, when providing sterilization of all 18 load types. Class B Sterilizers are class 1 medical device, extensively used in dentistry, medical and veterinary practices ensuring the safety of doctors, their staff and patients and legally protects their practice. A Class B Autoclave must meet the stringent European Standard EN 13060, which classifies loads into three categories based on their structure of the items to be sterilized such as hollow, porous and wrapped items. These systems work on the principle of pre and post vacuum technology, where a vacuum pump is fitted in the unit for air removal.
Bionics Scientific is setting up new standards in dental industry with its range of user-friendly Class B Autoclaves. Following European standards, these sterilizers are designed for providing efficient class b sterilization for all types of instruments and materials including all wrapped (single and double) or unwrapped items. Our standard models of class B sterilizers are available with volume 16 to 23 liters and offers performance guarantee of unmatched quality of Class B Cycles. These systems are easy to operate and do not require any training before operating.
| Adopts European Class B standard | Excellent vacuum dry effect |
| Automatic water filling option | Built-in steam generator |
| High quality stainless steel chamber | Meets European Standard EN 13060 |
| Stainless steel chamber | Digital control panel |
| Descriptive indictors and alarms | Independent Pressure Gauge |
| Built-in printer | PC connection interface |
| Over heat protection | Double locking system ensures safety |
| Model | BST/DA-16 | BST/DA-18 | BST/DA-23 | |
|---|---|---|---|---|
| Sterilization Class | European Class B standard Dental Autoclave | |||
| Capacity | 16 liters | 18 liters | 23 liter | |
| Sterilizing temperature | 121°C, 134°C | |||
| Special function | Kill the AIDS (HIV), hepatitis B (HBV) mad cow virus and Bacillus | |||
| Dry procedure | Vacuum drying | |||
| Display | LED screen | |||
| Test program | Bowie & Dick test, Helix test, Vacuum test | |||
| Safety Features | Safe Valve Double-control lock system and self-test alarm system | |||
| Water supply system | Two water tanks inside | |||
| Capacity of water tank | 3.5L clean water tank and 5 liter waste water tank | |||
| Instrument container | 5-layer tray rack with 3 trays | |||
| Voltage | 220V±10%, 50Hz | |||
| Optional Accessories | -Print by optional external mini-printer or download from USB port ( Sterilizing
data) -Extra stainless steel trays |
|||
| Program | Temperature | Sterilziation Time(Min) | Sterilziation Time(Min) | Total Time(Min) |
|---|---|---|---|---|
| SOLID | 134°C | 2.1 | 4 | 18-30 |
| SOLID | 121°C | 1.1 | 20 | 30-45 |
| WRAPPED | 134°C | 2.1 | 6 | 30-40 |
| WRAPPED | 121°C | 1.1 | 20 | 35-50 |
| TEXTILE | 134°C | 2.1 | 10 | 45-65 |
| TEXTILE | 121°C | 1.1 | 30 | 50-75 |
| PRION | 134°C | 2.1 | 18 | 45-70 |
| PLASTIC | 105°C | 0.2 | 30 | 50-65 |
| B&D TEST | 134°C | 2.1 | 3.5 | 22-35 |
| VACUUM TEST | ------ | ----- | 3.5 | 4-10 |
A Class B dental autoclave is an advanced steam sterilizer that uses a vacuum system to remove air before sterilization. It is suitable for sterilizing wrapped, hollow, and solid dental instruments.
A Class B autoclave ensures complete sterilization of dental instruments, helping clinics maintain infection control standards and patient safety.
A dental autoclave sterilizes instruments by using high-pressure saturated steam at high temperatures to destroy bacteria, viruses, fungi, and spores.
Class B autoclaves can sterilize dental handpieces, surgical instruments, mirrors, forceps, scalers, wrapped instruments, and hollow devices.
Most Class B dental autoclaves complete a sterilization cycle in approximately 30 to 45 minutes, including drying time.
Dental autoclaves typically operate at 121°C or 134°C, with 134°C being commonly used for rapid and effective sterilization.
Yes, a Class B dental autoclave is specifically designed to sterilize wrapped instruments, allowing them to remain sterile until use.
A biological spore test should be performed weekly, while professional preventive maintenance is recommended at least once a year.
Common causes include overloading the chamber, improper instrument arrangement, insufficient water, blocked filters, or damaged door seals.
The ideal size depends on patient volume and instrument load. An 18-litre or 23-litre Class B autoclave is suitable for most dental clinics and hospitals.
Class B Dental autoclave sterilizers manufactured by us are successfully working in many Government hospitals, research labs and universities. These dental autoclave machines have been appreciated by many industry professionals due to their simple, practical, and reliable Class B sterilization features. If you are looking to buy a Class B autoclave for your dental clinic, please contact our sales team for a brochure, pricing details, and expert guidance in selecting the right model for your requirements.
Related Products: Autoclave Sterilizers, Glass Bead Sterilizer