A Class II biosafety cabinet is a ventilated, HEPA-filtered enclosure that protects the operator, the work and the laboratory when you handle biological material. Air is drawn in at the front to protect you, filtered air flows down over the work to protect your samples, and the air leaving the cabinet passes through a HEPA filter to protect the room.
Bionics Scientific is a Biosafety Cabinet Class 2 manufacturer in India, offering Type A2, B1 and B2 cabinets for microbiological, pharmaceutical, clinical and research laboratories. We have been manufacturing laboratory and cleanroom equipment in India for 26 years.
Biosafety Cabinet Class 2 is designed to protect the operator, product and laboratory environment during microbiological, pharmaceutical, clinical and research applications. These Biosafety Cabinets are used to contain low-to-moderate risk biohazardous materials. They have downward airflow and HEPA filters that re-circulate air, providing the required level of protection from microorganisms and aerosols. They are widely used in drug preparation, chemotherapy preparation, clinical research, medical and pharmaceutical work, and life science and industrial laboratories, so the cabinet you choose must meet the standards that apply to your application.
A biosafety cabinet goes by several names in the research industry. Microbiological safety cabinet, biological safety cabinet (BSC) and biosafety hood are the common ones, and microbiologists use them interchangeably. A biosafety hood is not the same thing as a chemical fume hood, which is covered in the comparison below.
These protections only work if the cabinet is used correctly and its airflow is certified, so the sections on use and certification below matter as much as the cabinet itself.
To meet varying research and clinical needs, biosafety cabinets are designed mostly in three types: Type A2, Type B1 and Type B2. When a cabinet is used correctly, all three protect the operator, the sample and the environment, and they are tested against the same pass and fail criteria. The CDC and NIH publish BMBL biosafety guidance that laboratories use to choose and use these cabinets.
Each biosafety cabinet is differentiated by the amount of air re-circulation, how air is exhausted, and the pressure arrangement. Before buying one, you should know the basic differences between these three types:
| Point | Type A2 | Type B1 | Type B2 |
|---|---|---|---|
| Air recirculated in the workspace | About 70% | About 30–40% | None |
| Air exhausted | About 30% | About 60–70% | 100% |
| Best suited to | Most microbiology and cell culture work without volatile chemicals | Work with small amounts of volatile chemicals, where a ducted cabinet is needed | Work involving volatile toxic chemicals |
| Exhaust arrangement | Can be connected to a duct or released through a HEPA filter, depending on installation | Ducted to the outside | Ducted to the outside |
Chemical use is the main reason to move away from Type A2. Check the BMBL guidance and your risk assessment before you decide, and tell us what you plan to handle.
| Class | What it protects | How it works |
|---|---|---|
| Class I | Operator and environment, but not the sample | Room air is drawn in across the work, and exhaust air is HEPA filtered |
| Class II | Operator, sample and environment | Inflow at the front, HEPA-filtered downflow over the work, HEPA-filtered exhaust |
| Class III | Highest level of containment | A gas-tight enclosure, worked through attached gloves, for the most dangerous agents |
This page covers Class 2 cabinets. If you need a Class I or Class III cabinet, contact us with your requirement.
Class II cabinets are commonly used for work with agents at biosafety levels 1, 2 and 3. Your own risk assessment and the BMBL guidance decide the containment you need, not the cabinet alone.
These three are often confused, but they do different jobs.
| Equipment | Protects | Use it for |
|---|---|---|
| Biosafety cabinet (Class II) | Operator, sample and environment | Biological material, including infectious agents |
| Laminar air flow cabinet | The sample only | Clean work with non-hazardous material. A horizontal unit blows air toward the user, so it is not for infectious material |
| Laboratory fume hood | The operator from chemical fumes | Chemical work. It has no HEPA filtration to protect the product |
Bionics Scientific manufactures biosafety cabinets that are built to follow the European Standard EN 12469 and NSF standards, and NSF/ANSI 49 certification is listed as an option. Please confirm the certification for your model in the quotation. Our cabinets are available in four sizes (3 ft, 4 ft, 5 ft and 6 ft), as a ducted or recirculating type, and there are design options and accessories to suit most applications. Each cabinet is factory tested before it ships.
| Class 5 conditions (class 100) per ISO 14644-1 and 2 | 10° sloped front for comfortable working |
| Produces less heat and energy efficient | Highly efficient airflow management system |
| Lighting through fluorescent lamp | Filters can be changed easily from the front side |
| Durable construction for long service life | Durable wall plenum design |
| Easy to clean and maintain | Powder coated MS exterior |
| Low noise operation | Audio / visual sash alarm |
| Nominal Size | 2 ft. | 3 ft. | 4 ft. | 5 ft. | 6 ft. |
| Working area (ft.) | 2 x 2 x 2 | 3 x 2 x 2 | 4 x 2 x 2 | 5 x 2 x 2 | 6 x 2 x 2 |
| External size (ft.) | 2 x 3 x 7 | 3 x 3 x 7 | 4 x 3 x 7 | 5 x 3 x 7 | 6 x 3 x 7 |
| Inflow Velocity | 105 fpm (.53 m/s) |
| Downflow Velocity | 60 fpm (0.30 m/s) |
| Noise Level | <65 dba |
| Filter | HEPA filter 99.999% efficient (ULPA - Optional) |
| Particle retention | 0.3 microns |
| Light | Fluorescent light |
| Light Intensity | 90 - 120 foot-candles |
| Controller | Microprocessor controller w/ digital display |
| Body Construction | Wooden / PCRC Sheet / SS 304 / SS 316 |
| Table Construction | Stainless Steel 304 |
| Front Door | Frameless auto sliding glass door |
| Pressure Gauge | Magnehelic gauge |
| Blower assembly | 1/3 HP, Single Phase, 1440 RPM motor, Capacity 1000 CFM, Pressure 30 mm WG |
| Exhaust assembly | 300 CFM, ducting by PVC pipe with rain guard - Dia. 150mm |
| Certifications | NSF/ANSI 49 (Optional) |
| Optional | - Raised airflow grill - Spare UV lamp - Vacuum Tap - Exhaust (LH / RH / Top) |
![]() |
![]() |
![]() |
| Stainless Steel Casters | Telescoping Base Stand | Magnehelic Gauge |
![]() |
![]() |
![]() |
| Thermal Anemometer | Service Fixtures | UV Lamp |
A biosafety cabinet only protects you while its airflow and filters are performing as designed. To keep it that way, follow a routine cleaning and inspection schedule, and have the cabinet tested by a qualified technician at installation, after it is moved, after filter changes or major repairs, and at least once a year.
Between certifications, check the airflow, the surface, the motor, the lighting and the electrical switches. Watch the pressure gauge for signs that the filters are loading, and arrange for replacement when they are saturated. Filters should be changed only by a trained person, because contaminated filters need safe handling.
The price depends on the cabinet type, size, body material, exhaust arrangement, accessories and certification. Send us your requirement and we will quote for that configuration.
We are a manufacturer, not a reseller, and we have been making laboratory and cleanroom equipment in India for 26 years. We help you choose the cabinet type and size from what you will handle, we confirm the specification and certification for your model in writing, and each cabinet is factory tested before it ships. You can read more about Bionics Scientific or contact us directly.
Get the exact price, specifications and delivery timeline for Biosafety Cabinet Class 2 Type A2, B1 and B2. Our experts will help you choose the right model based on your lab requirement.
Free consultation, custom sizes from 3ft to 6ft and fast delivery across India.
Submit your enquiry now to get a quotation within 24 hours.
A Class II biosafety cabinet is a ventilated, HEPA-filtered enclosure that protects the operator, the sample and the laboratory environment when working with biological material.
They differ in how much air is recirculated and exhausted. Type A2 recirculates about 70 percent of the air, Type B1 exhausts about 60 to 70 percent, and Type B2 exhausts 100 percent with no recirculation.
Type A2 suits most microbiology and cell culture work. Choose Type B2 if you work with volatile toxic chemicals. Type B1 sits between the two. Check your risk assessment before deciding.
A biosafety cabinet protects the operator, the sample and the environment. A laminar air flow cabinet only protects the sample, and a horizontal one blows air toward the user, so it is not for infectious material.
Class II cabinets are used for work with agents at biosafety levels 1, 2 and 3. Your risk assessment and the CDC and NIH BMBL guidance decide the right level of containment.
Cabinets are available in 3, 4, 5 and 6 ft sizes. A 2 ft size is also listed in the dimensions table.
Inflow velocity is 105 fpm (0.53 m/s) and downflow velocity is 60 fpm (0.30 m/s).
NSF/ANSI 49 certification is listed as an option. Confirm the certification for your model in the quotation.
At installation, after it is moved, after filter changes or major repairs, and at least once a year, by a qualified technician.
Send the cabinet type, size, the materials you will handle, body material, exhaust arrangement, accessories, certification needs, power supply and delivery location.